Total Polyphenols Quantification Assay Kit

Q Colorimetric Beverages, Food Quantitative Uncoated
产品编号 ABIN7384588
发货至: 中国
  • 检测方法
    Colorimetric
    检测范围
    3.5-25 μg/mL
    最低检测浓度
    3.5 μg/mL
    应用范围
    Quantification (Q)
    原理
    The kit is designed for the rapid determination of polyphenols in various samples.
    样品类型
    Beverages, Food
    Analytical Method
    Quantitative
    特异性
    Specific for phenol content determination.
    产品特性
    Phenolic Quantification Assay is based on Folin-Ciocalteu method. The FC reagent contains phosphomolybdic/ phosphotungstic acid complexes1. The method relies on the transfer of electrons in alkaline medium from phenolic compounds to form a blue chromophor
    组件
    Plate and the reagents neccesary to perform the assay.
    试剂未包括
    Pipettes, reaction tubes, plate reader.
  • 说明

    1 ml (microassay)/ 20μL (microplate)

    实验时间
    15 min
    板类型
    Uncoated
    试剂准备

    Reagent A -FC reagent-needs to be diluted (1:10)

    实验流程

    MICROASSAY: 1.Pipette 1 mL of each standard, unknown sample or diluted sample replicate into separate clean test tubes. Refer to the Table 1 (see booklet) as a guide for diluting the standard phenol. For the diluent, use the same buffer as in the samples. 2.Add 1 mL of Reagent A -Folin-Ciocalteu Reagent previously diluted (1:10) in deionized water- to each tube. 3.Add 1 mL of Reagent B (Alkaline Working Solution) to each tube and measure the absorbance at 700 nm immediately. 4.Measure the absorbance of these standards, blanks and unknown samples at 700 nm. MICROPLATE: 1. Prepare standards containing a range of 1 to 10 μg phenol (Gallic acid) to a volume of 200 μL. Pipette 20 μL of each standard and unknown sample or diluted sample replicate into a microplate well. Refer to the Table 2 (see booklet) as a guide for diluting the standard phenol. For the diluent, use the same buffer as in the samples. 2.To each well, add 100 μL of Reagent A -FC reagent-previously diluted (1:10) in deionized water. 3.Add 80 μL ofReagent B (Alkaline Working Solution) to each well and measure the absorbance at 700 nm on a plate reader immediately. 4.Measure the absorbance of these standards, blanks and unknown samples at 700 nm.

    结果分析

    1.If the spectrophotometer or microplate reader was not zeroed with the blank, then substract the average blank value from the standard and unknown sample values. 2.Create a standard curve by plotting A700nm (y-axis) vs standard, μg (x-axis). Determine the unknown sample concentration using the standard curve. 3.Standard curve example for microplate assay procedure is shown in Figure 2 (see images). 4.Dilute the unknown samples until they reach an absorbance within the limits of the standard curve.

    限制
    仅限研究用
  • 储存液
    Sodium azide
    注意事项
    This product contains Sodium azide: a POISONOUS AND HAZARDOUS SUBSTANCE which should be handled by trained staff only.
    储存条件
    RT/4 °C
    储存方法
    FC Reagent A: RT, FC Reagent B: RT, FC Standard: 4 °C