电话:
+86 (0512) 65829739
传真:
+86 (010) 6788 5057
电子邮件:
orders@antibodies-online.cn

CRP ELISA 试剂盒

The Cow CRP ELISA Kit (ABIN955836) is a Colorimetric ELISA Kit designed to quantify Cow CRP.
产品编号 ABIN955836
发货至: 中国
Contact our Customer Service for availability and price in your country. Contact Info

Our Local Distributor

中国
北京 101111
No. 88 KeChuang 6th Street
Beijing Economic Technological Development Area
Room 801-803
4A Biotech Co.,Ltd.
Tel +86 (0512) 65829739 传真 +86 (010) 6788 5057

Quick Overview for CRP ELISA 试剂盒 (ABIN955836)

抗原

See all CRP ELISA试剂盒
CRP (C-Reactive Protein (CRP))

适用

  • 29
  • 18
  • 18
  • 10
  • 8
  • 7
  • 7
  • 6
  • 5
  • 3
  • 3
  • 2
  • 1
  • 1
  • 1
  • 1
Cow

检测方法

Colorimetric

实验类型

Sandwich ELISA

检测范围

46.88-3000 pg/mL

应用范围

ELISA
  • 最低检测浓度

    46.88 pg/mL

    原理

    The kit is a sandwich enzyme immunoassay for the in vitro quantitative measurement of bovine CRP in serum, plasma and other biological fluids.

    Analytical Method

    Quantitative

    特异性

    This assay has high sensitivity and excellent specificity for detection of bovine CRP. No significant cross- reactivity or interference between bovine CRP and analogues was observed. Note: Limited by current skills and knowledge, it is impossible for us - complete the cross-reactivity detection between bovine CRP and all the analogues, therefore, cross reaction may still exist.

    灵敏度

    The minimum detectable dose of bovine CRP is typically less than 22.15 pg/mL.
    The Sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest protein concentration that could be differentiated from zero. It was determined the mean O.D. Value of 20 replicates of the zero calibrator plus three standard deviations.

    组件

    Pre-coated, ready to use 96-well strip plate (1x)
    Calibrator (lyophilized) (2x)
    Calibrator Diluent (1 x 20 mL)
    Detection Reagent A (1 x 120 µL)
    Detection Reagent B (1 x 120 µL)
    Assay Diluent A (2X concentrate) (1 x 6 mL)
    Assay Diluent B (2X concentrate) (1 x 6 mL)
    TMB Substrate (1 x 9 mL)
    Stop Solution (1 x 6 mL)
    Wash Buffer (30X concentrate) (1 x 20 mL)
    Plate sealer for 96 wells (4x).

    试剂未包括

    1. Microplate reader with 450 +/- 10 nm filter.
    2. Precision single or multi-channel pipettes and disposable tips.
    3. Eppendorf Tubes for diluting samples.
    4. De-ionized or distilled water.
    5. Absorbent paper for blotting the microtiter plate.
    6. Container for Wash Solution.
  • 说明

    The calibration curve concentrations used for the ELISA's were 3, 000 pg/mL, 1500 pg/mL, 750 pg/mL, 375 pg/mL, 187.5 pg/mL, 93.75 pg/mL, 46.88 pg/mL.

    板类型

    Pre-coated

    结果分析

    Average the duplicate readings for each calibrator, control, and samples and subtract the average zero calibrator optical density. Create a calibration curve on log-log graph paper, with CRP concentration on the y-axis and absorbance on the x-axis. Draw the best fit straight line through the calibrator points and it can be determined by regression analysis. Using some plot software is also recommended. If samples have been diluted, the concentration read from the calibration curve must be multiplied by the dilution factor.

    限制

    仅限研究用
  • 储存条件

    -20 °C

    储存方法

    All the reagents should be kept according to the labels on vials. The Calibrator, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20° C upon being received. The unused strips should be kept in a sealed bag with the desiccant provided to minimize exposure to damp air. Opened test kits will remain stable until the expiration date shown, provided it is stored as prescribed above.

    有效期

    The expiry date is stated on the label.
  • 抗原 See all CRP ELISA试剂盒

    CRP (C-Reactive Protein (CRP))

    背景

    The microtiter plate provided in this kit has been pre-coated with a monoclonal antibody specific to CRP. Calibrators or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated polyclonal antibody preparation specific for CRP. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then a TMB substrate solution is added to each well. Only those wells that contain CRP, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of a sulfuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm +/- 10 nm. The concentration of CRP in the samples is then determined by comparing the O.D. of the samples to the calibration curve.