Hexokinase 1 ELISA 试剂盒

HK1 适用: 人 Colorimetric Sandwich ELISA 78-5000 pg/mL Cell Culture Supernatant, Plasma, Serum
产品编号 ABIN455345
发货至: 中国
  • 抗原 See all Hexokinase 1 (HK1) ELISA试剂盒
    Hexokinase 1 (HK1)
    适用
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    检测方法
    Colorimetric
    实验类型
    Sandwich ELISA
    检测范围
    78-5000 pg/mL
    最低检测浓度
    78 pg/mL
    应用范围
    ELISA
    原理
    This immunoassay kit allows for the specific measurement of human Hexokinase concentrations in cell culture supernates, serum and plasma.
    样品类型
    Cell Culture Supernatant, Serum, Plasma
    Analytical Method
    Quantitative
    特异性
    This assay recognizes recombinant and natural Human Hexokinase.
    产品特性
    Homo sapiens,Human,Hexokinase-1,Brain form hexokinase,Hexokinase type I,HK I,HK1,2.7.1.1
    组件
    Reagent (Quantity): Assay plate (1), Standard (2), Sample Diluent (1x20ml), Assay Diluent A (1x10ml), Assay Diluent B (1x10ml), Detection Reagent A (1x120µl), 2 Detection Reagent B (1x120µl), Wash Buffer(25 x concentrate) (1x30ml), Substrate (1x10ml), Stop Solution (1x10ml)
    Featured
    Discover our best selling HK1 ELISA Kit
  • 样本量
    100 μL
    板类型
    Pre-coated
    实验流程
    This assay employs the quantitative sandwich enzyme immunoassay technique. A monoclonal antibody specific for Hexokinase has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and any Hexokinase present is bound by the immobilized antibody. An enzyme-linked polyclonal antibody specific for Hexokinase is added to the wells. Following a wash to remove any unbound antibody-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Hexokinase bound in the initial step. The color development is stopped and the intensity of the color is measured.
    试剂准备

    Bring all reagents to room temperature before use. Wash Buffer - If crystals have formed in the concentrate, warm to room temperature and mix gently until the crystals have completely dissolved. Dilute 20 mL of Wash Buffer Concentrate into deionized or distilled water to prepare 500 ml of Wash Buffer. Standard - Reconstitute the Standard with 1.0 mL of Sample Diluent. This reconstitution produces a stock solution of 10000 pg/ml. Allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making serial dilutions. The undiluted standard serves as the high standard (10000 pg/ml). The Sample Diluent serves as the zero standard (0 pg/ml). Detection Reagent A and B - Dilute to the working concentration specified on the vial label using Assay Diluent A and B (1:100), respectively.

    样品收集
    Cell culture supernates - Remove particulates by centrifugation and assay immediately or aliquot and store samples at ≤ -20 °C. Avoid repeated freeze-thaw cycles. Serum - Use a serum separator tube (SST) and allow samples to clot for 30 minutes before centrifugation for 15 minutes at approximately 1000 x g. Remove serum and assay immediately or aliquot and store samples at -20 °C. Plasma - Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 15 minutes at 1000 x g at 2 - 8 °C within 30 minutes of collection. Store samples at ≤ -20 °C. Avoid repeated freeze-thaw cycles. Note: Citrate plasma has not been validated for use in this assay.
    实验流程

    Allow all reagents to reach room temperature. Arrange and label required number of strips.
    1. Prepare all reagents, working standards and samples as directed in the previous sections. 3
    2. Add 100 uL of Standard, Control, or sample per well. Cover with the adhesive strip. Incubate for 2 hours at 37 °C.
    3. Remove the liquid of each well, don’t wash.
    4. Add 100 uL of Detection Reagent A to each well. Incubate for 1 hour at 37°C. Detection Reagent A may appear cloudy. Warm to room temperature and mix gently until solution appears uniform.
    5. Aspirate each well and wash, repeating the process three times for a total of three washes. Wash by filling each well with Wash Buffer (350 uL) using a squirt bottle, multi-channel pipette, manifold dispenser or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot it against clean paper towels.
    6. Add 100 uL of Detection Reagent B to each well. Cover with a new adhesive strip.Incubate for 1 hours at 37 °C.
    7. Repeat the aspiration/wash as in step
    5. 8. Add 90 uL of Substrate Solution to each well. Incubate for 30 minutes at room temperature. Protect from light.
    9. Add 50 uL of Stop Solution to each well. If color change does not appear uniform, gently tap the plate to ensure thorough mixing.
    10. Determine the optical density of each well within 30 minutes, using a microplate reader set to 450 nm.
    Important Note:
    1. The wash procedure is critical. Insufficient washing will result in poor precision and falsely elevated absorbance readings.
    2. It is recommended that no more than 32 wells be used for each assay run if manual pipetting is used since pipetting of all standards, specimens and controls should be completed within 5 minutes. A full plate of 96 wells may be used if automated pipetting is available.
    3. Duplication of all standards and specimens, although not required, is recommended.
    4. When mixing or reconstituting protein solutions, always avoid foaming.
    5. To avoid cross-contamination, change pipette tips between additions of each standard level, between sample additions, and between reagent additions. Also, use separate reservoirs for each reagent.
    6. To ensure accurate results, proper adhesion of plate sealers during incubation steps is necessary.

    结果分析

    Average the duplicate readings for each standard, control, and sample and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a standard curve by plotting the mean absorbance for each standard on the y-axis against the concentration on the x-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the Hexokinase concentrations versus the log of the O.D. 4 and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.

    限制
    仅限研究用
  • 注意事项
    1. The kit should not be used beyond the expiration date on the kit label.
    2. Do not mix or substitute reagents with those from other lots or sources.
    3. If samples generate values higher than the highest standard, further dilute the samples with the Assay Diluent and repeat the assay. Any variation in standard diluent, operator, pipetting technique, washing technique,incubation time or temperature, and kit age can cause variation in binding.
    4. This assay is designed to eliminate interference by soluble receptors, ligands, binding proteins, and other factors present in biological samples. Until all factors have been tested in the Immunoassay, the possibility of interference cannot be excluded.
    储存条件
    4 °C/-20 °C
    储存方法
    The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20 °C upon being received. The other reagents can be stored at 4 °C.
  • 抗原 See all Hexokinase 1 (HK1) ELISA试剂盒
    Hexokinase 1 (HK1)
    别名
    HK1 (HK1 产品)
    别名
    HK1-ta ELISA Kit, HK1-tb ELISA Kit, HK1-tc ELISA Kit, HKD ELISA Kit, HKI ELISA Kit, HMSNR ELISA Kit, HXK1 ELISA Kit, Hexokinase ELISA Kit, hk1 ELISA Kit, HEX ELISA Kit, HEX-1 ELISA Kit, HK ELISA Kit, ARABIDOPSIS THALIANA HEXOKINASE 1 ELISA Kit, ATHXK1 ELISA Kit, F19B15.160 ELISA Kit, F19B15_160 ELISA Kit, GIN2 ELISA Kit, GLUCOSE INSENSITIVE 2 ELISA Kit, HEXOKINASE ELISA Kit, hexokinase 1 ELISA Kit, im:7148527 ELISA Kit, wu:fc09d08 ELISA Kit, wu:fc16e02 ELISA Kit, wu:fc21e02 ELISA Kit, wu:fq14b11 ELISA Kit, zgc:55790 ELISA Kit, zgc:77618 ELISA Kit, hexokinase-1 ELISA Kit, HK1 ELISA Kit, BB404130 ELISA Kit, Hk-1 ELISA Kit, Hk1-s ELISA Kit, dea ELISA Kit, mHk1-s ELISA Kit, HEXOKIN ELISA Kit, Hxk1 ELISA Kit, NtHxK1 ELISA Kit, NtHxk ELISA Kit, hexokinase ELISA Kit, StHK1 ELISA Kit, hxk ELISA Kit, hk1-A ELISA Kit, hexokinase 1 ELISA Kit, hexokinase-1 ELISA Kit, Hexokinase 1 ELISA Kit, hexokinase 1, putative ELISA Kit, hexokinase 1 L homeolog ELISA Kit, HK1 ELISA Kit, hk1 ELISA Kit, LOC100136279 ELISA Kit, Hex1 ELISA Kit, HXK1 ELISA Kit, Hk1 ELISA Kit, TA19800 ELISA Kit, hxk1 ELISA Kit, LOC107797523 ELISA Kit, LOC102604144 ELISA Kit, hk1.L ELISA Kit
    背景
    A hexokinase is an enzyme that phosphorylates a six-carbon sugar, a hexose, to a hexose phosphate. In most tissues and organisms, glucose is the most important substrate of hexokinases, and glucose-6-phosphate the most important product.Hexokinases have been found in every organism checked, ranging from bacteria, yeast, and plants to humans and other vertebrates. They are categorized as actin fold proteins, sharing a common ATP binding site core surrounded by more variable sequences that determine substrate affinities and other properties. Several hexokinase isoforms or isozymes providing different functions can occur in a single species. There are four important mammalian hexokinase isozymes (EC 2.7.1.1) that vary somewhat in their subcellular locations, kinetic characteristics with respect to different substrates and operating conditions, and physiological function. They are designated hexokinases I, II, III, and IV or hexokinases A, B, C, and D.
    途径
    Carbohydrate Homeostasis, Warburg Effect
You are here:
客服