Estradiol ELISA 试剂盒

适用: Various Species Colorimetric Competition ELISA 12.35 pg/mL - 1000 pg/mL Cell Culture Supernatant, Cell Lysate, Plasma, Serum, Tissue Homogenate
产品编号 ABIN415564
发货至: 中国
  • 抗原 See all Estradiol ELISA试剂盒
    Estradiol
    适用
    • 4
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Various Species
    检测方法
    Colorimetric
    实验类型
    Competition ELISA
    检测范围
    12.35 pg/mL - 1000 pg/mL
    最低检测浓度
    12.35 pg/mL
    应用范围
    ELISA
    原理
    The kit is a competitive inhibition enzyme immunoassay technique for the in vitro quantitative measurement of E2 in serum, plasma, tissue homogenates, cell lysates, cell culture supernates.

    We offer validation data (WB) for the kit components. So you can be sure to order a reliable ELISA kit product composed of high quality reagents.
    样品类型
    Cell Culture Supernatant, Cell Lysate, Plasma, Serum, Tissue Homogenate
    Analytical Method
    Quantitative
    特异性

    This assay has high sensitivity and excellent specificity for detection of Estradiol (E2).
    No significant cross-reactivity or interference between Estradiol (E2) and analogues was observed.

    交叉反应 (详细)
    No significant cross-reactivity or interference between Estradiol (E2) and analogues was observed.
    灵敏度
    4.93 pg/mL
    组件
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Standard Diluent
    • Detection Reagent A
    • Detection Reagent B
    • Assay Diluent A
    • Assay Diluent B
    • Reagent Diluent (if Detection Reagent is lyophilized)
    • TMB Substrate
    • Stop Solution
    • Wash Buffer (30 x concentrate)
    • Instruction manual
    Featured
    Discover our best selling Estradiol ELISA Kit
    Top Product
    Discover our top product Estradiol ELISA Kit
  • 应用备注
    • Limited by the current condition and scientific technology, we cannot completely conduct the comprehensive identification and analysis on the raw material provided by suppliers. So there might be some qualitative and technical risks to use the kit.
    • The final experimental results will be closely related to validity of the products, operation skills of the end users and the experimental environments. Please make sure that sufficient samples are available.
    • Kits from different batches may be a little different in detection range, sensitivity and color developing time.
    • Do not mix or substitute reagents from one kit lot to another. Use only the reagents supplied by manufacturer.
    • Protect all reagents from strong light during storage and incubation. All the bottle caps of reagents should be covered tightly to prevent the evaporation and contamination of microorganism.
    • There may be some foggy substance in the wells when the plate is opened at the first time. It will not have any effect on the final assay results. Do not remove microtiter plate from the storage bag until needed.
    • Wrong operations during the reagents preparation and loading, as well as incorrect parameter setting for the plate reader may lead to incorrect results. A microplate plate reader with a bandwidth of 10nm or less and an optical density range of 0-3 O.D. or greater at 450 ± 10nm wavelength is acceptable for use in absorbance measurement. Please read the instruction carefully and adjust the instrument prior to the experiment.
    • Even the same operator might get different results in two separate experiments. In order to get better reproducible results, the operation of every step in the assay should be controlled. Furthermore, a preliminary experiment before assay for each batch is recommended.
    • Each kit has been strictly passed Q.C test. However, results from end users might be inconsistent with our in-house data due to some unexpected transportation conditions or different lab equipments. Intra-assay variance among kits from different batches might arise from above factors, too.
    • Kits from different manufacturers for the same item might produce different results, since we have not compared our products with other manufacturers.
    说明

    Information on standard material:
    The standard might be recombinant protein or natural protein, that will depend on the specific kit. Moreover, the expression system is E.coli or yeast or mammal cell. There is 0.05% proclin 300 in the standard as preservative.

    Information on reagents:
    The stop solution used in the kit is sulfuric acid with concentration of 1 mol/L. And the wash solution is TBS. The standard diluent contains 0.02 % sodium azide, assay diluent A and assay diluent B contain 0.01% sodium azide. Some kits can contain is BSA in them.

    Information on antibodies:
    The provided antibodies and their host vary in different kits.

    样本量
    50 μL
    实验时间
    2 h
    板类型
    Pre-coated
    实验流程
    1. Prepare all reagents, samples and standards,
    2. Add 50μL standard or sample to each well.
      Then add 50μL prepared Detection Reagent A immediately.
      Shake and mix. Incubate 1 hour at 37 °C,
    3. Aspirate and wash 3 times,
    4. Add 100μL prepared Detection Reagent B. Incubate 30 minutes at 37 °C,
    5. Aspirate and wash 5 times,
    6. Add 90μL Substrate Solution. Incubate 10-20 minutes at 37 °C,
    7. Add 50μL Stop Solution. Read at 450 nm immediately.
    试剂准备
    1. Bring all kit components and samples to room temperature (18-25 °C) before use. If the kit will not be used up in one time, please only take out strips and reagents for present experiment, and leave the remaining strips and reagents in required condition.
    2. Standard - Reconstitute the Standard with 2.0mL of Standard Diluent, kept for 10 minutes at room temperature, shake gently (not to foam). The concentration of the standard in the stock solution is1,000pg/mL. Prepare 5 tubes containing 0.6mL Standard Diluent and produce a triple dilution serie. Mix each tube thoroughly before the next transfer. Set up 5 points of diluted standard such as 1,000pg/mL, 333.33pg/mL, 111.11pg/mL, 37.04pg/mL, 12.35pg/mL, and the last tube with Standard Diluent is the blank as 0pg/mL.
    3. Detection Reagent A and Detection Reagent B - If lyophilized reconstitute the Detection Reagent A with 150μL of Reagent Diluent, kept for 10 minutes at room temperature, shake gently (not to foam). Briefly spin or centrifuge the stock Detection A and Detection B before use. Dilute them to the working concentration 100-fold with Assay Diluent A and B, respectively.
    4. Wash Solution - Dilute 20 mL of Wash Solution concentrate (30x) with 580 mL of deionized or distilled water to prepare 600 mL of Wash Solution (1x).
    5. TMB substrate - Aspirate the needed dosage of the solution with sterilized tips and do not dump the residual solution into the vial again.

    Note:

    1. Making serial dilution in the wells directly is not permitted.
    2. Prepare standard within 15 minutes before assay. Please do not dissolve the reagents at 37 °C directly.
    3. Detection Reagent A and B are sticky solutions, therefore, slowly pipette them to reduce the volume errors.
    4. Please carefully reconstitute Standards or working Detection Reagent A and B according to the instruction, and avoid foaming and mix gently until the crystals are completely dissolved. To minimize imprecision caused by pipetting, use small volumes and ensure that pipettors are calibrated. It is recommended to suck more than 10μL for one pipetting.
    5. The reconstituted Standards, Detection Reagent A and Detection Reagent B can be used only once.
    6. If crystals have formed in the Wash Solution concentrate (30x), warm to room temperature and mix gently until the crystals are completely dissolved.
    7. Contaminated water or container for reagent preparation will influence the detection result.
    实验精密度

    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Estradiol (E2) were tested 20 times on one plate, respectively.
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Estradiol (E2) were tested on 3 different plates, 8 replicates in each plate.
    CV(%) = SD/meanX100
    Intra-Assay: CV<10%
    Inter-Assay: CV<12%

    限制
    仅限研究用
  • 注意事项
    The Stop Solution suggested for use with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
    注意事项
    The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage condition.
    To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
    储存条件
    4 °C
    储存方法
    • For unopened kit: All the reagents should be kept according to the labels on vials. The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20°C upon receipt while the others should be at 4°C.
    • For opened kit: When the kit is opened, the remaining reagents still need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and reseal along entire edge of zip-seal.
      Note: It is highly recommended to use the remaining reagents within 1 month provided this is within the expiration date of the kit.
    • For ELISA kit, 1 day storage at 37°C can be considered as 2 months at 4°C, which means 3 days at 37°C equaling 6 months at 4°C.
    有效期
    6 months
  • Huang, Liu, Jiang, Liao, Wei, Bi, Liu, Deng, Jin, Tan, Yang, Qin: "G-CSF-mobilized peripheral blood mononuclear cells combined with platelet-rich plasma accelerate restoration of ovarian function in cyclophosphamide-induced POI rats†." in: Biology of reproduction, Vol. 101, Issue 1, pp. 91-101, (2020) (PubMed).

    Zhou, Shi, Wang, Tan, Zhang: "Aqueous extract of lily bulb ameliorates menopause-like behavior in ovariectomized mice with novel brain-uterus mechanisms distinct from estrogen therapy." in: Biomedicine & pharmacotherapy, Vol. 117, pp. 109114, (2020) (PubMed).

    Guo, Xia, Zhang, Ma, Yan, Hao, Han, Ma, Zhou, Du et al.: "Menstrual blood derived mesenchymal stem cells combined with Bushen Tiaochong recipe improved chemotherapy-induced premature ovarian failure in mice by inhibiting GADD45b expression in the cell cycle ..." in: Reproductive biology and endocrinology : RB&E, Vol. 17, Issue 1, pp. 56, (2020) (PubMed).

    Wang, Liu, Xiong, Di, Yuan, Wu, Chen: "Reduced serotonin impairs long-term depression in basolateral amygdala complex and causes anxiety-like behaviors in a mouse model of perimenopause." in: Experimental neurology, Vol. 321, pp. 113030, (2020) (PubMed).

    Zhou, Zheng, Sharma, Zhang: "Total Polysaccharides of Lily Bulb Ameliorate Menopause-Like Behavior in Ovariectomized Mice: Multiple Mechanisms Distinct from Estrogen Therapy." in: Oxidative medicine and cellular longevity, Vol. 2019, pp. 6869350, (2020) (PubMed).

    Zhao, Ge, Wei, Liu, Liu, Ma, Zhao, Wei, Ma: "Effect of FSH on E2/GPR30-mediated mouse oocyte maturation in vitro." in: Cellular signalling, Vol. 66, pp. 109464, (2020) (PubMed).

    Wang, Zheng, Cui, Jiang, Zhou, Zhang, Hao: "Baicalin ameliorates polycystic ovary syndrome through AMP-activated protein kinase." in: Journal of ovarian research, Vol. 12, Issue 1, pp. 109, (2020) (PubMed).

    Carrillo, Collado, Díaz, Chowen, Grassi, Pinos: "Blocking of Estradiol Receptors ERα, ERβ and GPER During Development, Differentially Alters Energy Metabolism in Male and Female Rats." in: Neuroscience, Vol. 426, pp. 59-68, (2020) (PubMed).

    Cheng, Shen, Ding, Liu, Chu, Ma, Hou, Hao, Wang, Hou, Bai: "Lepidiline A Improves the Balance of Endogenous Sex Hormones and Increases Fecundity by Targeting HSD17B1." in: Molecular nutrition & food research, Vol. 64, Issue 10, pp. e1900706, (2020) (PubMed).

    Cavalcante, Ferreira, Pereira, Moraes, Simas, Sannomiya, Breithaupt-Faloppa, Moreira: "Acute administration of oestradiol or progesterone in a spinal cord ischaemia-reperfusion model in rats." in: Interactive cardiovascular and thoracic surgery, Vol. 26, Issue 2, pp. 196-201, (2019) (PubMed).

    Hassan, Desoky, Sakhr, Gabra, Bakri: "Possible Metabolic Alterations among Autistic Male Children: Clinical and Biochemical Approaches." in: Journal of molecular neuroscience : MN, Vol. 67, Issue 2, pp. 204-216, (2019) (PubMed).

    Chang, Han, Zhang, Zhou, Feng, Chen, Li, Li, Si: "G protein-coupled estrogen receptor activation improves contractile and diastolic functions in rat renal interlobular artery to protect against renal ischemia reperfusion injury." in: Biomedicine & pharmacotherapy, Vol. 112, pp. 108666, (2019) (PubMed).

    Oswari, Hidayat, Fatmawati, Hayati, Alisa: "Gambir Extract (Uncaria Gambir) Decreases Inflammatory Response and Increases Gastric Mucosal Integrity in Wistar Rats - Model Gastritis." in: Open access Macedonian journal of medical sciences, Vol. 7, Issue 19, pp. 3149-3152, (2019) (PubMed).

    Ding, Gong, Li, Fu, Jin, Zhang, Gao, Sun: "Rcan2 and estradiol independently regulate body weight in female mice." in: Oncotarget, Vol. 8, Issue 29, pp. 48098-48109, (2018) (PubMed).

    Wei, Song, Zhifen: "Molecular mechanism and functional role of macrophage colony‑stimulating factor in follicular granulosa cells." in: Molecular medicine reports, Vol. 16, Issue 3, pp. 2875-2880, (2018) (PubMed).

    Mao, Fan, Yu, Luo, Wu, Tang, Kang, Tang: "Abnormality of Klotho Signaling Is Involved in Polycystic Ovary Syndrome." in: Reproductive sciences (Thousand Oaks, Calif.), Vol. 25, Issue 3, pp. 372-383, (2018) (PubMed).

    Ding, Tan, Song, Ma, Xiao, Zhang: "Effect of controlled ovarian hyperstimulation on puberty and estrus in mice offspring." in: Reproduction (Cambridge, England), Vol. 154, Issue 4, pp. 433-444, (2018) (PubMed).

    Hua, Cao, Wang, Sun, Chen: "Exposure of Pregnant Mice to Triclosan Causes Insulin Resistance via Thyroxine Reduction." in: Toxicological sciences : an official journal of the Society of Toxicology, Vol. 160, Issue 1, pp. 150-160, (2018) (PubMed).

    Zhao, Zhou, Chen, Liu, Chu, Zhang: "Beneficial effects of Heqi san on rat model of polycystic ovary syndrome through the PI3K/AKT pathway." in: Daru : journal of Faculty of Pharmacy, Tehran University of Medical Sciences, Vol. 25, Issue 1, pp. 21, (2018) (PubMed).

    Wang, Wang, Lu, Cui, Li, Li, Zhang, Zhang, Liu: "The Combination of icariin and constrained dynamic loading stimulation attenuates bone loss in ovariectomy-induced osteoporotic mice." in: Journal of orthopaedic research : official publication of the Orthopaedic Research Society, Vol. 36, Issue 5, pp. 1415-1424, (2018) (PubMed).

  • 抗原 See all Estradiol ELISA试剂盒
    Estradiol
    Abstract
    Estradiol 产品
You are here:
客服